D-2 Ultrasound Gel Heater
D.3
Specifications
Power supply
Voltage
10-20V
(
±5%
)
Power consumption
12W±10%W
Environmental conditions
Operating conditions
Storage and transportation conditions
Ambient temperature
0°C~40°C
-20°C~55°C
Relative humidity
30%~85% (no condensation)
30%~95% (no condensation)
Atmospheric pressure
700hPa~1060hPa
700hPa~1060hPa
D.4
Function and Requirement
The gel heater can make the gel reach 40
℃
. There are four levels for the temperature: 34
℃
, 37
℃
, 40
℃
and off.
When the gel is placed inside the heater, the time it takes to heat from an ambient temperature of 18°C
to the desired temperature should be no more than 0.5 hours.
The ultrasound gel heater can work continuously over 12 hours.
NOTE:
1.
When the ambient temperature is higher than the required temperature of the
heater, the heater does not function.
2.
The heater can only heat one bottle of gel at a time.
D.5
Install the Heater
1. Push the gel heater into the gel box as shown in the figure.
3.
Indicator
The indicator is off when switching off the
heater.
Set the heater temperature to low; one
indicator on the right becomes white.
Set the heater temperature to med; two
indicators on the right become white.
Set the heater temperature to high; three
indicators on the right become white.
4.
Power supply indicator
The power supply indicator becomes green
after the power supply accessing.
5.
Power supply cable
Connect to the power socket under the control
panel.
6.
Bottom cover
/