5-114 Image Optimization
5.16.2 Screen Display of Tissue Tracking QA
1
—Displays image used to generate trace curve
2
—Displays ECG trace
3
—Displays measurement and calculation results:
EDV
:
Maximum value of the end diastolic volume during the trace.
EDA
:
Maximum value of the end diastolic area (Left Ventricular) during the trace.
ESV
:
Maximum value of the end systolic volume (Left Ventricular) during the trace.
ESA
:
Maximum value of the end systolic area (Left Ventricular) during the trace.
FAC (for short axis section): Fractional Area Change= (EDA
— ESA)/EDA
EF (for long axis section): Ejection fraction
HR
:
Heart rate
Global strain of all segments. Displays when strain rate curve is acquired.
Global strain rate of all segments. Displays when strain rate curve rate is acquired.
Also on Bull’s Eye figure, the system displays TPSD value:
Time to Peak Standard Deviation (TPSD):
Where, standardized value of time to peak data:
N
i
TP
i
,
1
|
. (N is the number of time to peak data)
Average value of standardized value of time to peak data:
TP
, and the standard deviation is
N
TP
TP
TPSD
N
i
1
2
.
4
—Display curve: Velocity/Displacement/Strain/Strain Rate.
Each curve on the image is matched with a certain segment in the cardiac segmentation model (6),
identified by different colors.
Velocity curve
X-axis represents time (s);
Y-axis represents velocity (cm/s).
1
2
3
4
5
6