5-84 Image Optimization
Tint and Tint map
Press the knob under [Tint Map] to activate the function, and then change the effect by rotating the
knob.
Ruler
Tap [Ruler] on the touch screen to hide or show the ruler around the image.
Color Display (in Power iScape)
Tap [Color Disp.] on the touch screen to display or hide the color image.
Color Map (in Power iScape)
Rotate the knob under [Color Map] on the touch screen to select the map, as in Power mode.
5.12.2.2 Image Zooming
Rotate <Zoom> to zoom in or zoom out the image.
Rotate the button clockwise to zoom in the image, the max factor is 10;
Rotate the button anticlockwise to zoom out the image.
Roll the trackball to change position of the magnified image.
Press <Zoom> again to exit zoom mode.
When the image displayed is bigger than the image area, the thumbnail is automatically
displayed.
5.12.2.3 Rotating the Image
For the convenience of image viewing, you can rotate the image by rotating the knob under [Rotation] on
the touch screen in increments of 5°.
5.12.2.4 Measurement, comments, and body mark
In iScape image review status, you can perform measurement, comment, and body mark.
NOTE:
1. The measurement accuracy for the sliced image may be degraded, exercise caution
when measurements are performed from an iScape image.
2. If there is a trace during the retracing, do not perform measurement across the trace.
5.12.2.5 Evaluate image quality
Many variables may affect the overall image quality. It is important to evaluate the image content and
quality before an image is used for diagnosis or measurements.
NOTE:
iScape panoramic imaging is intended for use by well-trained ultrasound operators or
physicians. The operator must recognize image items that will produce a sub-optimal or
unreliable image.
If the image quality cannot satisfy the following criteria, you shall remove the image and capture it again.
The Image must be continuous (no part of an image moves suddenly or disappears.)
No shadow or absent signal along the scan plane.
Clear profile of anatomy through the entire scan plane without distortion.
Skin line is continuous.
The images are captured from the same plane.
There are no large black areas in the image.