Image Optimization 5-73
5.11.13 Smart FLC (Smart Follicles Calculation)
CAUTION:
Smart FLC result is provided for reference only, not for confirming a diagnosis.
NOTE: Smart FLC is an optional function, and it is not available in Smart 3D mode.
5.11.13.1 Basic Procedures for Smart FLC
1. Acquire necessary 3D/4D data and freeze.
2. Tap
[Smart FLC] tab on the touch screen to enter Smart FLC, and the system is in the “Edit ROI”
status ([Edit ROI] button is highlighted in green);
3. Set the position and size of the ROI box.
4. Tap [Calc] on the touch screen.
5. Auto calculation is finished:
Section image (A, B, C) is enclosed with colored lines and the color corresponds to the follicle in
VR window.
3D image (VR);
Results on the right of the screen are displayed in a table.
V: Volume
d1, d2, d3: diameters of the follicle
MD: Follicle Diam
NOTE:
To ensure the correctness of the results, please select an image with clear follicle boundary
when entering smart FLC.
5.11.13.2 Operation Controls
Edit ROI
Same as the operation in ROI editing in Smart-V.
Edit/ Undo
Tap [Edit] on touch screen to turn on the editing function. It supports dividing, merging, adding and
deleting of the follicle.