DICOM/HL7 11-13
HL7 service setting for Worklist is described as follows:
Name
Note
Configure the New
service
Device Name
After you set the server (s) in DICOM Server Setting
screen, the name (s) will appear in the drop-down list,
select the name of the Worklist server.
DICOM Service
Name
Default is server-HL7Query, and it can be modified.
AE Title
Application Entity title.here, it should be consistent with
that of the HL7 server.
Port
DICOM communication port, 104 by default.Here, the port
should be consistent with that of the HL7 server port.
Maximum Retries
Reserved feature.
Interval Time(s)
Reserved feature.
Timeout
Refers to time after which the system will stop trying to
establish a connection to the service.Value: 5
-
60s, in
increments of 5, and 15 by default.
Monitoring mode
The ultrasound device starts using the monitoring port to
receive the date after the monitoring mode is enabled.
In monitoring mode, the ultrasound system sends the
information as the user end; it monitors the result via the
port as the service end.
In non-monitoring mode, the ultrasound system sends the
query information and receives the query result as the
user end; the monitoring mode is not chosen by default.
Monitoring port
It is the port to receiving the data for the ultrasound system
after the monitoring mode is enabled. Here, the port
should be consistent with that of the HL7 server port.
See the server settings for the monitoring port.
Add
Click to add the Worklist service to the service list.