5-12 Image Optimization
Dynamic Range
Description
Adjusts contrast resolution of an image, compresses or expands gray display range.The
real-time dynamic range value will be displayed on the image parameter area on the top
of the screen.
Operation
Turn the knob under [Dynamic Range] on the touch screen to complete the adjustment.
The adjusting range is 30-180 in increments of 5.
Impacts
The more the dynamic range is, the more specified the information appears.
M Soften
Description
This feature is used to process the scan lines of M images to reject noise, making the
image details to be clearer.
Operation
Tap [M Soften] to adjust the parameters on the touch screen.
The system provides 4 levels of M Soften adjustment. 0 represents the function is
disabled. The bigger the value is, the stronger the effect becomes.
5.4
Color Mode Image Optimization
The Color mode is used to detect color flow, and the color is designed to judge the direction and speed
of blood flow.
Generally, the color above the color bar indicates the flow towards the probe, while the color below the
color bar indicates the flow away from the probe; the brighter the color is, the faster the flow speed
becomes; while the darker the color is, the slower the flow speed becomes.
5.4.1
Color Mode Exam Protocol
1. Select a premium image during B mode scan, and adjust to place the area of interest in the center of
the image.
2. Press <C> to enter B+Color mode. Use the trackball and <Set> to change position and size of the
Region of Interest (ROI).
3. Adjust the image parameters during scan to obtain optimized images.
4. Perform other operations (e.g. measurement and calculation) if necessary.
5.4.2
Color Mode Image Optimization
In Color mode scan, the image parameter area on the right side of the screen displays the real-time
parameter values as follows:
Items
F
G
WF
PRF
Meaning Frequency Color Gain Color wall filter Pulse Repetition Frequency PRF