13-42 Probes and Biopsy
NGB-011
Hold the probe and the needle-guided bracket, and then open the grip knob of the needle-guided
bracket.
NGB-018
(1) Loosen the guiding block's nut and slightly move the guiding block in the direction of the
needle’s tail.
(2) Separate the residual part of the needle-guide bracket and the transducer from the needle.
(3) Loosen the bracket's pinch nut and remove the needle-guided bracket from the transducer.
NGB-021
Hold the transducer in the left hand; unscrew the locking nut with the right hand to loose the
retaining clamp. Raise the needle-guided bracket to separate the locating clamp/ bulge from the
locating grooves.
NGB-022
(1) Pull the lock pin and open up the V-shaped cover to expose the needle.
(2) Separate the bracket and the transducer from the needle.
(3) Screw the pinch nut to release the needle-guided bracket.
(4) Separate the bracket and the transducer.
NGB-024
1. Hold the transducer. Press the biopsy needle to separate the needle from pressure position of the
needle.
2. Separate the bracket and the transducer from the needle.