Probes and Biopsy 13-35
3. Tighten the pinch nut of the needle-guided bracket to confirm that the needle-guided bracket is
properly installed on the transducer.
4. Adjust the dial scale to the required needle type shift.
5. Adjust the needle angle to the proper shift as required (loosen the nut first, and then tighten the nut
based on the shift you need).
6. Pull the lock pin and close the V-shaped cover to fix the lock pin in the groove of the needle type
adjusting base, so as to install the needle into the guiding hole.
Needle-guide Bracket Installation (NGB-036)
1. Put on the sterile probe sheath.
2. Select a proper needle-guided bracket, and match the tab with the groove of the transducer. Mount
the bracket onto the transducer.
3. Rotate the clamping nuts of the guided bracket on the right and left side to fix the bracket and the
transducer. Rotate the needle-type adjusting nut to the ultimate position as shown in the figure.
4. Hold the transducer. Press the biopsy needle pressure position to separate it from needle guided
V-shaped block.