Wipe
13.1.9 Storage and Transportation for Bracket (Fusion)
When all examinations for the day have been completed, confirm that the transducer bracket is in good
condition. After disinfecting the probe, confirm that the transducer bracket is in good condition and stored
in a suitable place.
1. To prevent the transducer bracket from being damaged, DO NOT store it in locations where it may be
exposed to:
Direct sunlight
Sudden changes in temperature
Dust
Excessive vibration
Heat generators
2. Store and transport under following ambient conditions:
Ambient temperature: -20°C ~ 55°C
Relative humidity: 20% to 95% (no condensation)
Atmospheric pressure: 700 hPa ~1060 hPa
13.1.10 Disposal of the Transducer Bracket
Be sure to dispose of the needle-guided bracket only after sterilizing it.
Contact your MINDRAY representative when disposing of this device.