Patient Data Management 10-3
The thumbnail of this image will appear in the thumbnail area on the right side of the screen. When
you move the cursor onto the thumbnail, its file name with suffix will be displayed.
Set the cine length via [Preset]
[Systen]
[Image].
10.2.5 Quickly Saving Full Screen Image to the System
Specific operation:
1. Set user-defined key: select one key from [Preset]
[System]
[Key Config] page. Select “Full
Screen” from “Output” column.
2. Press the user-defined key to save the image.The format of the image is PNG.
The full screen image is saved to the default path with the default name. The thumbnail of this image will
appear in the thumbnail area on the lower side of the screen. When you move the cursor onto the
thumbnail, its file name with suffix will be displayed.
10.2.6 Quickly Saving Images to USB Flash Drive
Use user-defined keys to quickly save the single-frame or cine to USB flash drive.
1. Set the user-defined key via the path: [Setup]
[System]
[Key Config].
Select a key
in the Key Function field on the left side, and select “Save Image to USB disk” in the
Output page of the Function field on the right side. Click [Save] to exit setup.
2. Press the user-defined key to save the image to the USB flash drive.
10.2.7 Quickly Exporting Cine File to USB Flash Drive
Use the user-defined key to export the cine to the USB flash drive (not saved to the local path).
The file is stored in the directory: X (USB flash drive name): patient name+ ID\exam type + exam time \
image ID.
1. Set the user-defined key through the path: [Setup]
[System Preset]
[Key Config]. Select a key in
the Key Function page on the left side and select “Send Cine to USB disk” in the Output page of
Function field on the right side. Click [Save] to exit.
2. Perform the scan and freeze the image
3.
Press the user-defined key to save the image to the USB flash drive.
10.2.8 Auxiliary Output Function
For the following three functions, the system provides auxiliary output function setting: “Save Image”,
"Save Clip
(Retrospective)" and “Save Clip (Prospective)”.When the corresponding user-defined key is
pressed, the ultrasound system can perform multiple operations one by one as per the preset.
Save the single-frame image:
Send image to DICOM storage
Send image to DICOM print
Send image to USB disk
Send image to iStorage
Send image to local printer
Retrospective/prospective saving cine:
Send cine to DICOM storage
Send cine to USB disk
Send cine to iStorage
Setting method (taking
“Save Image” as an example, add the auxiliary functions “Send Image to DICOM
Storage” and “Send Image to USB disk”):