5-88 Image Optimization
5.13.1.7 iTouch
On contrast status, you can also get a better image effect by using iTouch function.
Press <iTouch> on the control panel to turn on the function.
The symbol of iTouch will be displayed in the image parameter area in the upper right corner of the
screen once press <iTouch>.
Select different levels of iTouch effect through [iTouch] on the touch screen.
Long press <iTouch> to exit the function.
5.13.1.8 Image Saving
Live capture
In live mode, you can save the interested images by tapping [Pro Capture] and [Retro Capture].
Cine saving
In live mode, press <Freeze> on the control panel to enter cine review status.
5.13.2 Left Ventricular Opacification
Basic Procedures for LVO:
1. Acquire ECG signal
;
2. Tap <Probe> on the touch screen to open Probe/Exam Mode selecting dialogue box;
3. Select SP5-1U probe and LVO exam mode;
4.
Workflow of LVO is similar to abdomen contrast imaging. For details, please refer to “5.13.1 Basic
Procedures for Contrast Imaging
5.13.3 Measurement, Comment and Body Mark
The system supports image measurement, comment and body mark functions. For the details, please
refer to the relevant sections.
5.13.4 Micro Flow Enhancement
MFE superimposes and processes multiple frames of contrast image during the cycle; it indicates tiny
vessel structures in detail by recording and imaging microbubbles.
MFE switch
During real-time scanning, tap [MFE] on the touch screen to start MFE imaging (MFE key is highlighted
in green).
Tips: MFE imaging is available only for real-time imaging or cine file of auto review mode. If a MFE cine
is reviewed manually, MFE effect cannot be displayed.
MFE Period
Rotate the knob under [MFE Period] on the touch screen to select different imaging period suitable for
current flow. Where, MAX is the maximum superimposing effect.
NOTE:
In MFE status, patient should lie down and hold breath, and transducer should be kept still.