Probes and Biopsy 13-59
No.
Image
Description
5
Indicating
box
Image with mark: the box indicates the needle tip projecting
point to the mark.
Image with no mark: the box indicates the needle tip projecting
point to the intersection in out-plane status.
The distance indicating box shrinks by the time needle tip goes
towards the lesion mark or intersection. When the tip is 3mm away
from the lesion mark or intersection, the box turns green.
6
Needle end
projecting
point
Projecting point of the needle end on the ultrasound image plane.
13.3.6 Disposal
Dispose needle, sensor bracket, sterile cover by requirements of the manufacturer. Contact the manufacturer
for details.
Dispose the bracket after sterilization. Contact Mindray for bracket disposal.
13.4 Middle Line
Middle Line helps to locate and observe the focus point of lithotripsy wave during lithotripsy treatment.
By means of providing information for the lithotripsy machine as well as a tool for watching the procedure
of lithotripsy in real-time, you can adjust the intension and frequency of the lithotripsy wave through
lithotripsy machine.
To enter the mode: tap [Middle Line] in the biopsy tab or preset a shortcut key for middle line function.
The middle line is a vertical dotted line located in the middle of the screen, the position and
direction of which cannot be changed.
There is a mark icon of “×” located on the middle line which can be moved up and down along
the line by rolling the track ball.
To use the Middle Line function of the ultrasound system:
The depth of the mark is displayed in the image parameter area of the screen.