5-78 Image Optimization
NOTE
The automatic measurement results appear only for these having the same
characteristics on the planes.
3. Tap [Edit] to edit the measurements. The measurements on the right of the screen update in real
time.
a) Tap [Edit] to modify the measurements. The caliper becomes green.
Or, press <Set> to activate the caliper (becoming green).
b) Move the trackball and press <Update> to modify the length and the position of the caliper.
c) Press <Set> to confirm the caliper. The caliper becomes white. [Edit] is off.
NOTE:
Once the automatic measurements are completed, the operations to axial rotation, reference line
rotation, parameter adjustment, MSP editing, zooming/panning, dual/quad-split display will remove the
measurement results.
Reset
All planes
Acquire 3D data. Tap [S-Planes CNS] to enter the automatic detection of the mode. Reset the
following operations:
Tap [All Planes] to reset the planes. The position and the angle to TCP, TTP, TVP, MSP and
reference line return to the initial condition.
Tap [All Planes] to reset the planes. The position to TCP, TTP, TVP, MSP and reference line
returns to the last MSP condition. The angle to CP, TTP, TVP, MSP and reference line
returns to 0°.
Current plane
Acquire 3D data. Tap [S-Planes CNS] to enter the automatic detection of the mode. Reset the
following operations:
Tap [Current Plane] to reset the plane. The position of the current plane returns to the initial
condition. The angle of the current plane returns to 0°.
Tap [Current Plane] to reset the current plane. The position and the angle to the current
plane returns to the last MSP condition.