7-6 Physiological Unit Signal
7.5
Parameter Description
The physio parameters are described as follows:
Parameter
Description
ECG Source
Select ECG source.
Gain
Function: to set the amplitude of the trace.
Method: click [Gain] on the soft menu, or press/rotate the corresponding knob.
Each adjustment takes you to the next setting.
Value: 0-30, in increments of 1.
Position
Function: to set the vertical position of the both traces on the image display.
Method: click [Position] on the soft menu. Each adjustment cycles you to the next
setting.
Value: 0%-100%, in increments of 5%.
Speed
Function: change the speed of the physio trace.
Value: 10-90mm/s.
T1
Function: to set the delay time T1 in Single trigger or Dual trigger.
T2
Function: to set the delay time T2.
Method: select a proper value in the drop-down list.
Interval
Function: to set the time interval for Timer.
Method: select a proper value.
Invert
To invert the display.