System Preparation 3-9
3.5.3
Probe Adapter Installation
NOTE: P7-3Ts and L16-4Hs can be connected to the ultrasound system through probe adaper
installation.
1. Firstly, move the probe adapter locking lever to horizontal direction and keep TEE probe locking
lever to unlock position. Then, push probe adapter to probe port slowly and carefully while keeping
the socket position upward.
2. Move the probe adapter locking lever to vertical position through 90
°
clockwise and lock it.
3. Stick the TEE probe connector to TEE probe socket and keep the cable upward, then move the
lever to locking position in order to lock the TEE probe, achieving the installation of probe adapter
and TEE probe.
Loose
Probe adapter locking lever
TEE probe socket unlock position
TEE probe socket
TEE probe locking lever
Tighte
n
Probe adapter
locking lever