5-138 Image Optimization
3. Tap [Set Modeling ROI] to highlight it (it is activated by default). Move the trackball to set ROI
position and size.
Set ROI position and size. Envelope the liver with the ROI.
Press <Set> again to adjust ROI position and size, and to set the start frame and end frame of the
motion modeling.
4. Tap [Motion Modeling]. If modelling succeeds, the system will play the cineloop automatically, and
ROI moves along with the motion of the respiration curve.
Note: RMQF scale is 0~1. 0 represents the poor motion modeling; 1 represents the premium motion
modeling.
Conduct step 4~5 repeatedly based on your demands. Set motion modeling repeatedly until a
premium one appears.
5. Tap [Motion Compen] to activate it. Move the probe. The Ultrasound System shows the CT image
which is processed by respiration compensation (Fusion Imaging with the respiration
compensation).
6.
Save multi-frame cine.