Image Optimization 5-105
Tip: When you confirm the ROI size by pressing <Update>, you cannot adjust the ROI size during
acquisition. You can only adjust the ROI position using the trackball.
4. Press <Save> on the control panel to start acquisition.
The system displays the Protocol window on the left of the screen, listing the phases for the selected
protocol along with the first phase views (phases are stages). The system selects the first view for
acquisition by default, indicating the selected view by placing a red mark to the right of the view name.
5. Proceed through each view in each stage according to the following instructions:
Non-continuous stages:
To save acquired images for the selected view, press<Save> on the control panel. The system
goes to the next view for acquisition by default, saved views are marked with a green “√.”
Rotate the knob under [Stages XXX] or [Views XXX] to select the stage and view for image
acquisition (or reacquisition). Press<Save> to start acquisition.
Tip: Views can be re-acquired until you click [End Acquisition].
If the protocol contains continuous stages (for alternative workflows), then proceed through each
continuous stage according to the following instructions:
To begin saving acquired images for the selected stage, press <Save> on the control panel.
The system displays a percentage marker below the selected stage indicating the progress of the
continuous capture.
To halt saving acquired images for the selected stage, rotate the knob under [Pause] on the
touch screen or press <Freeze> directly. The percentage stops increasing.
Select [Continue] or press <Freeze> again to continue.
To end the current acquisition, press <Save> on the control panel.
To select another continuous stage, rotate the knob under [Stage XXX] on the touch screen.
Suspending is not allowed under continuous exam.
When acquisition is complete for each stage, the system advances to the next stage. If the stage is
non-continuous, the system displays the stage views. When image acquisition is completed for all
views and continuous stages, the system switch to Select Mode.
6. To start or restart the timer, click [Stage Timer]/[Exam Timer] on the soft menu to turn it on.
The Stage time is displayed to the right side of each stage in the protocol list, while the Exam time is
displayed in the left side of the screen.
Each saved image will be marked with two times T1 and T2. T1 refers to the total time of the whole
acquisition, while T2 indicates the time the acquisition lasted for a certain stage.
7. To review loops before ending acquisition, select [Review/WMS] on the menu. You can redisplay the
real-time imaging screen to continue acquisition by selecting [Acquire].
8. To end the acquisition and review the acquired images, click [End Acquisition].
NOTE:
When the acquisition is ended, no stress echo image acquisition can be performed for the
same exam.
5.15.3
Selecting Preferred Stress Echo Loops (Select Mode)
The selected clips are used for analysis in the review mode and wall motion scoring mode. Select Mode
is used to select the best loops of the examination.
When the acquisition is ended, select mode is enabled automatically.
In Select Mode, you can select the represe
ntative loop (“preferred” loop) for each view.
To select “preferred” loops:
1. Use <Update> according to tips in the trackball hint area or select [Select] on the menu to enter the
Select Mode, or the system enters the Select Mode directly after the acquisition is finished.
2. Select the loop. Use the buttons displayed during Select Mode to designate another loop or another
view for display.