Probes and Biopsy 13-45
NGB-034
1. Hold the transducer, and press the pressure position of the needle to separate the biopsy needle
from the biopsy needle pressure position.
2. Remove the transducer and the needle-guided bracket to separate the biopsy needle.
3. Rotate the clamping nuts of the guided bracket on the right and left side to release the bracket and
the transducer. Remove the bracket from the transducer.
13.2.8 Clean and Sterilize the Needle-guided Bracket
Cleaning
Please follow the instructions in the manual for cleaning.
1. Wear sterile gloves to prevent infection.
2. Wash the needle-guided bracket with water or soap water to remove all the external matters. Or,
clean the needle-guided bracket with urethane sponge.
3. Wipe off the water on the needle-guided bracket using sterile cloth or gauze after washing it.
Sterilization
1. Wear sterile gloves to prevent infection.