3-8 System Preparation
3.4
Control Panel Position Adjustment
1. Connect to the power supply of the ultrasound system.
2. Press the control lever
to rotate the control panel ±90°. If the moving assembly of the control
panel is purchased by the user, the control panel can be moved back and forth. Press
to
ascend or descend the control panel.
3.5
Connecting/Disconnecting a Probe
CAUTION:
1.
Press <Freeze> to freeze an image or turn off the power of the system
before connecting/disconnecting the probe.Otherwise, system or probe
failure may occur.
2.
When connecting or disconnecting a transducer, place it in a proper
position, to prevent the transducer from falling off or becoming damaged.
3.
Hang the probe cable to the hanger located under the control panel to avoid
excessively bending and damaging the cable.
4.
Only use the transducers provided by Mindray. Aftermarket transducers
may result in damage or cause a fire.
3.5.1
Connecting a Probe
WARNING:
The transducers, cables and connectors are in proper operating order
and free from surface defects, cracks and peeling. Otherwise, electrical
shock may result.
1. Turn the knob horizontally. Keep the cable end of the transducer upwards, and insert the connector
into the port of the system, and then press in fully. (Shown in the following figure)
2. Turn the lock handle 90° clockwise to lock it securely. (Shown as the right figure)
3. Place the probe properly to avoid being treaded or wrapping around other devices. DO NOT allow
the probe head to hang free.
3.5.2
Disconnecting a Probe
1. Turn the locking lever 90°counterclockwise to the horizontal position. (Shown in the left figure)
2. Pull the probe connector straight out vertically (Shown in the figure below).
Lock