Image Optimization 5-93
Where, the X-axis represents time and Y-axis represents pressure.
5.14.1.4 Parameter adjusting
ROI Adjustment
Description
To adjust the width and position of ROI in Elasto imaging.
Operation
When the ROI box is solid line, roll the trackball to change its position.
When the ROI box is dotted line, roll the trackball to change the size.
Press <Set> to switch between the solid line and the dotted line status.
Smooth
Description
To adjust the smooth feature of the Elasto image.
Operation
Tap [Smooth] on the touch screen to adjust the parameters.
The system provides 5 levels of smooth function. The bigger the value is, the higher the
smooth becomes.
Opacity
Description
To adjust the opacity feature of the Elasto image.
Operation
Turn the knob under [Opacity] on the touch screen to complete the adjustment.
The system provides 5 levels of opacity function: the bigger the value the stronger the
effect.
Invert
Description
To invert the E color bar and therefore invert the colors of benign and malignant tissue.
Operation
Tap [Invert] on the touch screen.
Display Format
Description
Adjust the display format of ultrasound image and the Elasto image.
Operation
Tap [H 1:1], [V 1:1], [Full] on the touch screen to adjust.
The system provides 3 types of display format:
H 1:1: right and left display (the real-time ultrasound image appears on the left, and the
elasto image appears on the right);
V 1:1: up down display (the elasto image appears above, and the real-time ultrasound
image appears below).
Full: the elasto image only displayed.
Impacts
Adjust according to the actual situation and obtain a desired analysis through
comparison.
Map
Description
To select different maps for observation.