5-122 Image Optimization
<3>
Magnetic
generator
Transmitter
port
Plug the magnetic generator to the Transmitter port. Place
it around the patient
’s shoulder or two sides of the patient’s
abdomen. The magnetic generator should be frontal-faced
with the lesion. The frontal face of the magnetic generator
keeps 15-60 cm from the target within the scope 30 cm
up/down.
NOTE: keep the magnetic generator away from the metal
devices or electromagnetic interference.
<4>~<8>
Sensor (2),
Positioning
(probe)
navigation
support
Motion
(abdomen)
sensor support
Positioning
(probe)
sensor port
Motion
(abdomen)
sensor port
1. Plug two sensors to positioning (probe) sensor port
, and motion (abdomen) sensor port
respectively.
2. Positioning (probe) navigation support: fix the support
to the probe. See the method below:
The plastic markers on the probe and on the
navigation support must be on the same side.
The silk prints on the probe and the navigation
support must be on the same side.
Push the navigation support fastened following
the arrow
’s guide.
Silk print on the
support
Silk print on the
probe
Marker on the support
Marker on the probe
Positioning
(probe)
sensor port
Motion
(abdomen)
sensor
port