Probes and Biopsy 13-7
13.1.4 Wearing the Probe Sheath
A legally marketed probe sheath must be installed over the probe before performing intra-cavitary and
intra-operative examination. Protective barriers may be required to minimize disease transmission.
Probe sheaths are available for use with all clinical situations where infection is a concern.
To order probe sheath, contact:
CIVCO Medical Instruments Co.
102 First Street South, Kalona, IA 52247-9589 USA
Tel: 1-319-656-4447
E-mail: [email protected]
CAUTION:
Be sure to cover the probe with a new (unused) probe sheath to prevent
infection during examination. If the package of a probe sheath is open or
broken, the sterilization of the probe sheath may not be sufficient. DO NOT
use such a probe sheath.
The cover contains natural rubber latex and talc that can cause allergic
reactions in some individuals.
DO NOT use an expired probe sheath. Before using a probe sheath, verify
whether the term of validity has expired.
Method (for reference only):
1. Place an appropriate amount of gel inside the
sheath or on the probe acoustic lens. Poor
imaging may result if no gel is used.
2. Insert the probe into the sheath; make sure to
use proper sterile technique. Pull cover
tightly over probe acoustic lens to remove
wrinkles and air bubbles, and taking care to
avoid puncturing the sheath.
3. Secure the sheath with the enclosed elastic
bands.
4. Inspect the sheath to ensure there is no hole
or tear.