Probes and Biopsy 13-41
Needle Steer
Description
This function adjusts the biopsy needle angle via changing the steer angle of the scan
line.The iNeedle affecting region changes correspondingly.
Operation
Tap [Needle Steer] to adjust the angel on the touch screen in increment of 10°.
B/iNeedle
Description
This function is used to display B image and iNeedle image synchronously.
Operation
To disable or enbale the function, tap [B/iNeedle] on the touch screen.
Tips: iZoom (full-screen magnification) is available on iNeedle status.
13.2.7 Removing the Needle-guided Bracket
NGB-007
Metal needle-guided bracket:
1. Screw the nut of the guiding block and remove the guiding block slightly along the direction of
the needle
’s tail.
2. Separate the residual part of the needle-guide bracket and the transducer from the needle.
3. Screw the pinch nut of the bracket, and remove the needle-guided bracket from the transducer.
4. Separate the transducer and the needle-guided bracket.
plastic/needle detachable
1. Remove the guiding block slightly along the direction of the needle
’s tail.
2. Separate the residual part of the needle-guide bracket and the transducer from the needle.
3. Remove the support of needle-guided bracket from the transducer.