1-4
Safety
Precautions
10.
Do not allow the patient to contact the live parts of the ultrasound
system or other devices, e.g. signal I/O ports. Electric shock may
occur.
11.
Do not use an aftermarket probe other than those specified by
Mindray. The probes may damage the system causing a profound
failure, e.g. a fire in the worst case.
12.
Do not subject the transducers to knocks or drops.Use of a
defective transducer may cause an electric shock.
13.
Do not open the covers and front panel of the system. Short circuit
or electric shock may result when the system hardware is exposed
and powered on.
14.
Do not use this system when any digital device such as a
high-frequency electrotome, high-frequency therapeutic device or
defibrillator is applied already.Otherwise, there is a risk of electric
shock to the patient.
15.
Only use the ECG leads provided with the physiology module;
otherwise, electric shock may be resulted.
16.
When moving the system, you should hold the handle; otherwise,
damage may be resulted by abnormal force.Do not push the system
from the left/right side, otherwise, it may be toppled over.
17.
The auxiliary power output outlet in the system is used to supply
power for the recommended peripheral devices.Do not connect
other devices to the outlet, otherwise, the rated output power may
be exceeded and failure may be resulted.Maximum output power of
the outlet is 300 VA.
18.
Accessory equipment (analog or digital) connected to the
ultrasound system must comply with the relevant IEC standards
(e.g., IEC 60950 information technology equipment safety standard
and IEC 60601-1 medical equipment standard). Furthermore, all
configurations must comply with the standard IEC60601-1 3rd
chapter 16. It is the responsibility of the person, who connects
additional equipment to the signal input or output ports and
configures a medical system, to verify that the system complies
with the requirements of IEC60601-1 3rd chapter 16. If you have any
questions regarding these requirements, consult your vendor.
19.
Prolonged and repeated use of keyboards may result in hand or
arm nerve disorders for some individuals. Observe the local safety
or health regulations concerning the use of keyboards.
20.
When using intra-cavity transducers, do not activate the transducer
outside the patient’s body.
21.
It is not allowed for the operator to have contact with other patients
and the electronic parts (such as the input/output terminal of the
signal) of other devices that are connected to the system.
Otherwise, it may produce the electrical shock to the patient.