13-28 Probes and Biopsy
(1) Put on the sterile probe sheath.
(2) Hold the probe by one hand, select the proper needle-guided bracket, and hold it with the other
hand. Align the narrow end tab of the needle-guided bracket with the groove of the probe, then
push the needle-guided bracket forward, making the tabs and the grooves of the needle-guided
bracket to match with the grooves and tabs of the probe.
(3) Check manually to confirm that the needle-guided bracket is securely installed on the probe.
(4) Select a proper guiding block and push it into the groove above the angle block, and clamp it
tightly.
(5) Insert a biopsy needle with the same specification as that of the guiding block into the hole of
the guiding block.
NGB-011
(1) Inosculate the locating groove on the clamp with the two raised edges on the probe head and
align the locating pit of the clamp to the convex point on the probe head.
(2) Tighten the grip knob of the biopsy.
(3) Pull the biopsy guide through the hole and tighten the knob..
NGB-018
1. Put on the transducer sheath.
2. Select a proper needle-guided bracket, and match the groove with the tab of the transducer
respectively. Mount the bracket onto the transducer. The needle-guided brackets may be different
from each other, but the methods are the same.
For NGB-003/NGB-016/NGB-018