Safety
Precautions
1-5
Caution:
1.
Precautions concerning clinical examination techniques:
This system must be used only by qualified medical professionals.
This operator’s manual does not describe clinical examination
techniques. The clinician should select the proper examination
techniques based on specialized training and clinical experience.
2.
Malfunctions due to radio wave:
If a radio wave emitting device is used in the proximity of this system, it
may interfere with operations. Do not use or take any devices
transmitting RF signals (such as cellular phones, transceivers and radio
controlled products) in the room placing the system.
If a person brings a device that generates radio waves near the system,
ask him / her to immediately turn OFF the device.
3.
Precautions concerning movement of the system:
Please install the system on a flat plane with casters locked. Otherwise,
damage may be resulted by accidental moving.
Do not move the system laterally, which may result in damage in case of
toppling.
Move the system slowly on the slope by two people, otherwise, damage
may result in case of unexpected sliding.
Do not sit on the system, which may result individual falling in case of
system moving.
Object placed on the monitor may fall and injure an individual.
Fasten and fully secure any peripheral device before moving the system.
A loose peripheral device may fall and injure an individual.
When move the system on the steps, please take care to prevent the
system from toppling.
4.
If the circuit protector is tripped, it indicates that the system or a peripheral
device was improperly shut down and the system is unstable. You cannot
repair the system under this circumstance and must call the Mindray
Customer Service Department or sales representative.
5.
There is no risk of high-temperature burns during normal ultrasound
examinations. It is possible for the surface temperature of the transducer to
exceed the body temperature of a patient due to environmental temperature
and exam type combinations. Do not apply the transducer to the same region
on the patient for a long time. Apply the transducer only for a period of time
required for the purpose of diagnosis.
6.
Do not use the system to examine a fetus for a long period of time.
7.
The system and its accessories are not disinfected or sterilized prior to
delivery. The operator is responsible for the cleaning and disinfection of
probes and sterilization of biopsy brackets according to the manuals, prior to
the use. All items must be thoroughly processed to completely remove
harmful residual chemicals, which will not only harmful to the human body,
but also damage the accessory.
8.
It is necessary to press <End Exam> to end the current scan that is in
progress and clear the current Patient Information field. Otherwise, new
patient data may be combined with the previous patient data.
9.
Do not connect or disconnect the system’s power cord or its accessories
(e.g., a printer or a recorder) without turning OFF the system power first. This
may damage the system and its accessories or cause electric shock.