Operation
98
Instructions for Use Fabius MRI SW 3.n
z
If the ventilator settings are correct, confirm the
mode change.
z
If the ventilator settings are not correct, for each
parameter that needs to change, press the cor-
responding soft key, select the correct value,
and confirm the change. When the parameter
changes are completed, confirm the ventilation
mode change.
z
After the mode change is confirmed, the
»
Volume Control
« key LED switches from
blinking to constantly on, the ventilator switches
to the Volume Control mode, and the waveform
is restored.
The parameters that can be set for Volume Control
mode are shown the adjacent table, along with their
adjustment ranges and factory default values.
502
Ventilation Parameter
(Volume Control
Mode)
Adjustment
Range
Factory
Default
Value
Pressure Limitation
P
MAX
[cmH
2
O] ([hPa])
15 to 70
min. PEEP+10
40
Tidal Volume
V
T
[mL]
20 to 1400
600
Frequency
Freq [bpm] ([1/min])
4 to 60
12
Insp time:
Exp time
T
I
:T
E
4:1 to 1:4
1:2
Insp pause time:
Insp time
T
IP
:T
I
[%]
0 to 50
10
PEEP [cmH
2
O] ([hPa]) 0 to 20
0
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...