
Instructions for Use Fabius MRI SW 3.n
59
Assembly
z
Connect a gas cylinder (
1
) to its yoke as speci-
fied below:
z
Remove the old washer (
2
) and install a new
washer on the seat of the yoke gas inlet con-
nection.
z
Verify that the two index pins (
3
) below the gas
inlet (
4
) are present.
z
Insert the head (
5
) of the gas cylinder into the
yoke from below. Ensure that the gas outlet and
indexing holes on the cylinder head align with
the gas inlet and index pins of the yoke assem-
bly (
6
).
z
Engage the indexing holes with the index pins.
z
Turn the yoke handle (
7
) clockwise against the
cylinder head, so that the point of the yoke han-
dle bolt is aligned with the indent on the back of
the cylinder head. Verify that the washer is in
place, the index pins are engaged, and the cyl-
inder hangs vertically.
z
Tighten the yoke firmly.
z
When required, the cylinder valve (
8
) is opened
using the cylinder wrench (
9
) that is provided.
z
When a cylinder is removed, place the yoke
plug (
10
) in the yoke assembly and tighten.
053
7
6
10
9
1
5
2
8
4
3
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...