Technical Data
Recommended separation distances
Information regarding separation distances
(IEC 60601-1-2: 2001, tables 205 and 206)
Recommended separation distances between portable and mobile RF telecommunication
devices and the Fabius MRI
max.
P
EIRP
(W)
3 V/m
distance
1)
(m)
1) 3 V/m distance to transmitters with frequencies from 150 kHz to 2.5 GHz, otherwise 1 V/m distance.
1 V/m
distance
1)
(m)
Note
0.001
0.06
0.17
0.003
0.10
0.30
0.010
0.18
0.55
0.030
0.32
0.95
e.g. WLAN 5250 / 5775 (Europe)
0.100
0.58
1.73
e.g. WLAN 2440 (Europe), Bluetooth
0.200
0.82
2.46
e.g. WLAN 5250 (not in Europe)
0.250
0.91
2.75
e.g. DECT devices
1.000
1.83
5.48
e.g. GSM 1800- / GSM 1900- / UMTS- mobiles,
WLAN 5600 (not in Europe)
2.000
2.60
7.78
e.g. GSM 900 mobiles
3.000
3.16
9.49
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...