Technical Data
212
Instructions for Use Fabius MRI SW 3.n
Classification
II b
Conforming to Directive 93/42/EEC Appendix IX
UMDNS Code
10-134
Universal Medical Device Nomenclature System
Control Inputs Ranges
Low Oxygen Supply Pressure Alarm
APL-Valve
MAN mode
5 to 70 cmH
2
O (hPa)
SPONT mode
1.5 cmH
2
O (hPa)
accuracy from 5 to 15 L/min
±15 % of set value or ±3 cmH
2
O (hPa) (greater
applies)
pressure drop at 30 L/min
3.4 cmH
2
O (hPa) (wet and dry)
Alarm limit
Warning signal when the pressure drops below 20 ±4 psi
(1.4 ±0.3 kPa x 100)
Alarm signal
High priority alarm (Warning)
LED indicator
The red LED indicator in the O
2
area of the gas flow control interface will
flash until the O
2
supply is restored.
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...