Instructions for Use Fabius MRI SW 3.n
197
Technical Data
Gas supply from supplementary O
2
and N
2
O cylinders
(with pin-index connections)
Cylinder Connections
Pin-indexed hanger yokes (CGA V-1-1994)
Cylinder Gas Pressure
(typical full loads at 70 ºF, 21 ºC)
O
2
, AIR
1900 psi (131 kPa x 100)
N
2
O
745 psi (51.3 kPa x 100)
Cylinder Gauges
Conform to ASME B40.1 Grade B
Cylinder Gauge Range
O
2
0 to 3000 psi (206.8 kPa x 100)
N
2
O
0 to 3000 psi (206.8 kPa x 100)
AIR
0 to 3000 psi (206.8 kPa x 100)
Compressed gas supply at workstation inlet
Dew point
>5 °C (41 °F) at ambient temperature
Oil content
<0.1 mg/m
3
Particles
dust-free air (filtered with pores <1 µm)
Internal Regulator Safety Relief Valve Pressure
70 psi (4.8 kPa x 100) opening pressure
Protection Class
I, in accordance with IEC 60601-1
Applied parts:
Breathing system (nozzles, breathing hoses)
Type BF
Ingress of Fluids
IPX0
Power supply
, Rating non-configurable
100 to 240 VAC, 50/60 Hz, 8.7 A including auxiliary
power sockets
Rechargeable batteries
Rating:
24 V; 3.5 Ah
Type:
sealed, gelled lead-acid
Recharging time:
≤
16 hours on the mains or full operation time
Operation time with fully charged batteries:
45 minutes, minimum
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...