Instructions for Use Fabius MRI SW 3.n
Cleaning
Removing anesthetic gas scavenging
system (AGSS)
z
Remove the anesthetic gas receiving system
(AGS) including scavenging hose and exhaust
hose from the medical device.
Removing the Suction System
z
Remove suction bottle assembly, including bot-
tle and regulator.
Reprocessing Breathing system
All parts of the respiratory system, ventilator
diaphragm, Y-piece, respiratory hoses, respira-
tory bag, parts of the absorber, parts of the sec-
retion suction unit and parts of the anesthetic
gas scavenging system.
z
Thermally disinfect in the cleaning/disinfection
machine at 200 °F (93 °C) /10 minutes.
Only use neutral or mild alcalic cleaning agents
(e.g. Neodisher Medizym, Neodisher FA) and
completely demineralized water!
When using thermal disinfection, the addition of
disinfectant chemicals is not necessary; risk of
corrosion!
WARNING
Risk of infection
Always wear protective gloves when emptying
the suction bottle.
Observe the hospital hygiene regulations.
NOTE
Reprocessing/disinfection instructions for the
reusable suction bottle and the suction regulator,
refer to their respective Instructions for Use.
WARNING
After washing, a hot steam sterilization is
required to completely dry the breathing sys-
tem.
Insufficient drying of the control areas found
in the valve plate can lead to impairment of
device function or to failure of the medical
device.
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...