Instructions for Use Fabius MRI SW 3.n
105
Operation
SIMV/PS Ventilation
Synchronized Intermittent Mandatory Ventilation
(SIMV) mode is a mixture of mechanical ventilation
and spontaneous breathing. In SIMV mode, the
patient can breathe spontaneously. SIMV will
attempt to synchronize the mandatory ventilation
strokes with spontaneous efforts.
The mandatory ventilation strokes are the same as
those for volume ventilation. They are defined by
the parameters
V
T
,
Freq
,
T
INSP
,
T
IP
: T
I
, and
PEEP
.
Pressure support can be added during SIMV mode
to augment the patient’s spontaneous breathing
efforts. Adjusting the
Δ
P
PS
level to a value other
than
OFF
will enable Pressure Support during
SIMV mode. (Refer to “Pressure Support Ventila-
tion” on page 102 for additional information on
Pressure Support ventilation.)
Starting SIMV/PS Ventilation
The following examples and illustrations describe
starting SIMV/PS ventilation from the present ven-
tilation mode Volume Control:
1
Press the »
SIMV/PS
« key. The LED associated
with this key starts blinking. It remains blinking
until the selected mode of operation is con-
firmed.
2
The Waveform window is replaced by the venti-
lator settings window and a message that pro-
vides instructions to confirm the mode change.
z
If the ventilator settings are correct, confirm the
mode change.
z
If the ventilator settings are not correct, for each
parameter that needs to change, press the cor-
responding soft key, select the correct value,
and confirm the change. When the parameter
changes are completed, confirm the ventilation
mode change.
565
2
1
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...