Operation Concept
44
Instructions for Use Fabius MRI SW 3.n
Selecting/Setting Ventilation Parameters
The following example describes changing the V
T
(tidal volume) parameter in Volume Control mode:
1
In Volume Control mode, press the »
Volume
Control
« key. The Volume Control Ventilation
Settings window replaces the Waveform win-
dow.
2
Press the »
V
T
« soft key. The key becomes high-
lighted.
z
Select a new value.
z
Confirm the new value.
NOTE
There is a 15-second timeout period for making
ventilation mode changes, with a 3-tone audible
sequence after the first 10 seconds. If the new set-
ting is not confirmed within the timeout period, the
current ventilation setting remains in effect and the
Ventilation Settings window returns to the Wave-
form window.
509
1
510
2
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...