Operation
104
Instructions for Use Fabius MRI SW 3.n
The parameters that can be set for Pressure Sup-
port mode are shown the adjacent table, along with
their adjustment ranges and factory default values.
Pressure Support (
PS
) can be used to support the
spontaneous breathing of the patient.
If the inspiration flow (
Insp Flow
) during inspiratory
effort is greater than the set trigger flow (
TRIG-
GER
), the device supports the patient according to
the Pressure Support setting (
Δ
P
PS
).
The set inspiratory flow (
Insp Flow
) defines how
fast the
Δ
P
PS
pressure is reached. The end of
inspiration is automatically triggered when reaching
25% of maximum inspiratory flow (
Insp Flow
) or
after a maximum of 4 s. The value
Freq Min
(e.g.,3 bpm (1/min)) is used to set a safety period
(safety period = 1/
Freq Min
, e.g., 20 s). If no inspi-
ratory effort is detected and the safety period has
expired, the device applies a pressure-controlled
breath with
P
INSP
=
Δ
P
PS
.
The lower pressure limit
P
AW
low
is used for pres-
sure monitoring to detect apneas (disconnection)
and continuous pressure. An alarm is generated
when the pressure curve does not cross the pres-
sure threshold either from above or below.
Ventilation Parameter
(Pressure Support
mode)
Adjustment
Range
Factory
Default
Value
Support pressure
Δ
P
PS
[cmH
2
O] ([hPa])
3 to 20, OFF 10
Min frequency for apnea
vent.
Freq Min [bpm] ([1/min])
3 to 20, OFF 3
Trigger Sensitivity
Trigger [L/min]
2 to 15
2
Inspiratory Flow
Insp Flow
[L/min]
10 to 85
30
PEEP [cmH
2
O] ([hPa])
0 to 20
0
111
1
1
Trigger
Trigger
Insp Flow
Freq Min
Freq Min
25%
25%
P
AW
Flow
max.
4 sec
Apnea
ventilation
Flow Trigger
Flow Trigger
No Trigger
Time
Time
Δ
P
PS
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...