Operation
94
Instructions for Use Fabius MRI SW 3.n
Ventilation
Manual/Spontaneous Ventilation Mode
ManSpont (Manual/Spontaneous) is a non-auto-
matic mode of ventilation. However, the ventilation
monitor and alarms are still operational. In
ManSpont mode, the ventilator piston is moved par-
tially upward to reduce system compliance. Manual
ventilation (with APL valve pressure limit) can be
delivered with the APL valve in the MAN position.
Spontaneous ventilation (APL valve wide-open)
can occur with the APL valve in the »
Spont
« posi-
tion.
The following examples and illustrations describe
starting Man/Spont ventilation from the present
ventilation mode »
Volume Control
«:
For Spontaneous Breathing:
z
Rotate the APL valve knob fully counterclock-
wise until the index mark on the knob lines up
with the index mark on the bottom of the
valve.The valve is now open for spontaneous
patient breathing.
z
Set the appropriate fresh gas flow.
1
Press the »
ManSpont
« key. The LED associ-
ated with this key starts blinking. It remains
blinking until the selected mode of operation is
confirmed.
2
The Waveform window is replaced by the
ManSpont window and a message that pro-
vides instructions to confirm the mode change.
WARNING
Risk of strangulation
Use caution when connecting the patient.
022
518
1
2
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...