Instructions for Use Fabius MRI SW 3.n
209
Technical Data
O
2
Cell Measurement Performance
Response time
Less than
16 seconds
measured values are not pressure
compensated
Warm-up time
after 5 min
error
≤
3 % of measured value
Drift Sensitivity
±1 % of measured value / 8h
Cross Sensitivity
≤
1 vol.% O
2
at 70 vol.% N
2
O and 5 vol.% CO
2
with 4 vol.% Halothane
or with 5 vol.% Enflurane
or with 5 vol.% Isoflurane
or with 10 vol.% Sevoflurane
effect of humidity / sensitivity
max. ±0.02 % of measured value / % relative humidity
Service life of O
2
sensor cell
>12 months at 25 °C (77 °F), 50 % relative humidity, 50 % O
2
gas mixture (or >5000 hours at 100 Vol% O
2
)
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...