Instructions for Use Fabius MRI SW 3.n
91
Operation
WARNING
Do not flush dry gas continuously for unne-
cessarily long periods through the soda lime
in the anesthesia system!
Otherwise the soda lime will dehydrate. If the
moisture level falls below a minimum level,
undesirable reactions generally occur, regard-
less of the type of soda lime and the inhalation
anesthetic used:
– reduced CO
2
absorption,
– increased heat generation in the absorber
and therefore increased breathing gas
temperature,
– CO formation,
– absorption and/or breakdown of the inha-
lationanesthetic.
The above mentioned reactions may endanger
the patient, e.g.:
– CO poisoning
– insufficient depth of anesthesia
– burns of the airway.
CAUTION
Soda lime irritates the skin and there is a risk of
serious damage to eyes.
If soda lime has leaked:
– Powdered soda lime must not be inhaled or
swallowed.
– Put on protective gloves and goggles, or a
face mask.
– In case of coming in contact with eyes, rinse
immediately with large amounts of water and
consult a physician immediately, otherwise it
may lead to eye damage.
– Powdered soda lime on the skin must be
washed off immediately, because it may irri-
tate the skin.
NOTE
Please refer to the specific Instructions for Use for
"Drägersorb 800 Plus or Drägersorb FREE".
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...