205
Technical Data
Insp Flow
Inspiratory flow
10 to 75 L/min (1 L/min resolution) in Pressure Control
mode
10 to 85 L/min (1 L/min resolution) in PS and SIMV/PS
modes
Δ
P
PS
(Pressure
Support)
Support Pressure
3 to 20 cmH
2
O (1 cmH
2
O resolution)
(3 to 20 hPa (1 hPa resolution))
Δ
P
PS
(SIMV/PS)
Support Pressure
3 to 20 cmH
2
O, OFF (1 cmH
2
O resolution)
(3 to 20 hPa, OFF (1 hPa resolution))
Freq Min
Apnea Ventilation minimum
frequency
3 to 20 bpm (1 bpm resolution) and “OFF”
(3 to 20 1/min (1/min resolution) and “OFF”)
Trigger
Trigger Level
2 to 15 L/min (1 L/min resolution)
T
INSP
SIMV Inspiratory Time
0.3 to 4.0 s
Pressure Support Ventilation Mode
The Pressure Support Ventilation mode has been verified under the following range of simulated patient
conditions:
Endotracheal Tube Size:
0.18 in to 0.32 in (4.5 mm to 8 mm)
Patient Lung Compliance:
10 mL/cmH
2
O to 100 mL/cmH
2
O
(10 mL/hPa to 100 mL/hPa)
Unassisted Patient Tidal Volume:
50 mL to 1000 mL
Patient Breath Rate (bpm) (1/min):
10 to 35
Delivery Accuracy
P
MAX
Pressure limiting
±5 cmH
2
O (±5 hPa) of setting
V
T
Tidal volume
±5 % of setting or 20 mL, whichever is greater
discharged to atmosphere, no compliance compensa-
tion)
f
Breathing frequency
±1 bpm (1/min) of setting
T
I
/T
E
Inspiration/expiration ratio
±5 % of setting
T
IP
/T
I
Inspiration pause
±25 % of setting
PEEP
End-expiratory pressure
±2 cmH
2
O (±2 hPa) or ±20 % of setting, whichever is
greater
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...