System Overview
24
Instructions for Use Fabius MRI SW 3.n
Auxiliary Oxygen Flowmeter
The auxiliary oxygen flowmeter delivers a metered
flow of pure oxygen, used, for example in the deliv-
ery of oxygen through a nasal cannula. Auxiliary
oxygen can be used in any ventilation mode, in
Standby, or even if the machine is switched off.
The auxiliary oxygen flowmeter may be used to
provide supplemental inspired oxygen to a patient
under spinal, epidural, or other regional anesthesia.
It may also be used to enrich the inspired gas mix-
ture provided by a manually powered selfinflating
resusciator bag*.
z
Test the auxiliary oxygen flowmeter. Adjust the
flow knob (
1
) and make sure the float moves
freely over the full range of the flowmeter.
When finishing oxygen therapy, make sure the flow-
meter is completely closed:
z
Turn the flow knob (
1
) clockwise until it can no
longer be turned.
Only then the oxygen flow is completely off.
*
ASTM F1850-22(2005) §76
WARNING
Risk of patient injury
Do not connect the patient directly to the auxi-
liary oxygen outlet.
High pressure will be applied and the patient
endangered.
WARNING
Risk of fire
Cauterizing close to a source of oxygen can
lead to fire.
WARNING
Risk of fire
Before cauterizing, close the flowmeter,
remove the mask and wait a few moments to
ensure that any oxygen accumulation has dis-
sipated.
069
1
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...