170
Instructions for Use Fabius MRI SW 3.n
Cleaning
Disinfecting components or accessories
z
Disinfect parts by immersing.
z
Sufficiently rinse parts under running water until
no disinfectant residues can be recognized.
z
Inspect parts for visible soiling and damage.
Repeat manual disinfection if necessary.
z
Shake out remaining water thoroughly. Allow
parts to dry thoroughly.
Visual inspection
z
Inspect all parts for damage and wear, e.g., cra-
cking, embrittlement or pronounced hardening
and residual soiling.
Sterilization
Use a vacuum steam sterilizer (in accordance with
DIN EN 285), preferably with fractional vacuum, for
sterilization.
z
Hot steam sterilization can be carried out at
134 °C (273.2 °F). Observe Instructions for
Use of medical device.
Reprocessing and disinfection
This section contains instructions for dismantling
and cleaning the Fabius MRI anesthesia work-
place.
During the reprocessing cycles, the vaporizers
remain attached to the medical device.
Device surfaces
CAUTION
Even accessories designed to be reused (e.g.
after reprocessing) have a limited service life. Due
to a number of factors connected with handling
and reprocessing (e.g.disinfectant residues can
attack the material more intensely during
auto-claving), increased wear can occur and the
service life can be markedly shortened. These
parts must be replaced if signs of wear become
visible, such as cracks, deformation, discolora-
tion, peeling, etc.
CAUTION
Do not sterilize parts in ethylene oxide! Ethylene
oxide may diffuse into the parts and cause
damage to health.
CAUTION
The Spirolog Flow Sensor and the Infinity ID Flow
Sensor must not be sterilized in hot steam. The
flow sensors are not resistant to high temperatu-
res and may be damaged.
CAUTION
When moving the writing tray, the holding arms
and drawers, keep a distance from the edges to
avoid crushing.
CAUTION
Risk of damage to the medical device
The surfaces of Fabius MRI, the pressure gas
hoses and cables must not be treated with alcohol
containing agents.
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...