Operation
114
Instructions for Use Fabius MRI SW 3.n
Preparing for Storage or Transport
WARNING
When moving the anesthesia work-
station, remove all monitors and
equipment from the top shelf and
hinged arms, remove the absorber
system, vaporizers, and reserve gas
cylinders, push in the writing tray.
The anesthesia workstation should only be
moved by people who are physically capable
of handling its weight. Dräger recommends
that two people move the anesthesia worksta-
tion to aid in maneuverability.
Exercise special care so that the machine
does not tip when moving up or down incli-
nes, around corners, and across thresholds
(for example, in door frames and elevators).
Do not attempt to pull the machine over any
hoses, cords, or other obstacles on the floor.
WARNING
All the procedures described in this manual
require that any use of ferromagnetic tools
can only be done while the Fabius MRI is out
of the scanner room.
WARNING
Do not bring any ferromagnetic tools or equip-
ment into the scanning room.
Ferromagnetic objects (made of steel, iron, or
stainless steel) are strongly attracted to the
magnet and can become harmful projectiles.
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...