Instructions for Use Fabius MRI SW 3.n
153
Configuration
Volume Alarms On/Off
1
Press the »
Volume Alarms On
« soft key in the
Setup window. The key label changes from
»
Volume Alarms On
«
to »
Volume Alarms
Off
«, and the volume alarms are disabled.
Auto Set
2
Press the »
Auto Set
« soft key in the Setup win-
dow. The breathing pressure threshold is set to
4 cmH
2
O (hPa) below the current plateau pres-
sure data value.
NOTE
The »
Volume Alarms On/Off
« soft key label does
not appear in ManSpont mode because it is select-
able on the ManSpont screen.
NOTE
The threshold setting may not be less than
5 cmH
2
O (5 hPa) or greater than 30 cmH
2
O
(30 hPa).
NOTE
In the absence of a current plateau pressure data
value, pressing the soft key will have no effect.
NOTE
In SIMV/PS mode, the breathing pressure thresh-
old will be set relative to the mandatory ventilation
stroke.
503
1
2
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...