Assembly
54
Instructions for Use Fabius MRI SW 3.n
Installing the Drägersorb CLIC
Adapter*
The disposable absorber:
– CLIC Absorber 800+,or
– CLIC Absorber Free, or
– Infinity ID CLIC Absorber 800+, or
– Infinity ID CLIC Absorber Free
can be used on the Fabius MRI by using the
Drägersorb CLIC adapter.
For information on installing the Drägersorb CLIC
adapter, consult its Instructions for Use.
To click the absorber into place:
1
Press the button; the mounting swings open.
z
Before fitting, shake the disposable absorber,
e.g. by turning it upside down several times in
order to loosen up the soda lime.
z
Remove seal from new disposable absorber.
z
Slide the new disposable absorber onto the
mounting and
2
Push the disposable absorber in the direction of
the machine until it engages.
*
optional
WARNING
The disposable absorber must be clicked into
place before switching on the Fabius MRI, so
that the absorber is included in the leak and
compliance test for the machine.
Otherwise breathing circuit leaks could be
undetected.
076
1
2
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...