Instructions for Use Fabius MRI SW 3.n
95
Operation
z
Confirm the mode change. The ManSpont
screen is activated. After the mode change is
confirmed, the »
ManSpont
« LED switches from
blinking to constantly on and the waveform is
restored.
The ManSpont screen allows the user to adjust two
parameters: Apnea Pressure alarm ON/OFF (see
page 129) and Volume alarms ON/OFF (see
page 127). Pressing the »
ON/OFF
« soft key tog-
gles the corresponding alarm between ON and
OFF.
For Manual Ventilation:
z
Rotate the APL valve adjustment knob to the
desired pressure. Clockwise rotation increases
the pressure, and counterclockwise rotation
decreases the pressure.
NOTE
In ManSpont mode, the apnea volume timer count-
down for caution alarms changes from 15 seconds
to 30 seconds, and for warning alarms from
30 seconds to 60 seconds.
517
023
20
10
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...