Fault-Cause-Remedy
166
Instructions for Use Fabius MRI SW 3.n
!!!
(25)
PRESSURE NEGATIVE
Measured P
AW
is
≤
-5 cmH
2
O
(hPa),
or MEAN
is
≤
-2 cmH
2
O (hPa).
Check breathing circuit and
ventilator settings.
!
(8)
PRESSURE SENSOR FAIL
Faulty sensor or pressure not
calibrated.
Call DrägerService or your
local authorized service orga-
nization.
!
(2)
PRES THRESHOLD LOW
Ventilation parameters were
changed without changing
alarm settings (see page 143).
Push the Auto Set soft key and
check ventilator settings.
!
(1)
RS232 COM1 FAIL
External monitor cable discon-
nected from External Commu-
nication Port 1.
Check monitor interface cable.
!
(1)
SPEAKER FAIL
Speaker failed.
Call DrägerService or your
local authorized service orga-
nization.
!!!
(28)
VENTILATOR FAIL
Ventilator not assembled cor-
rectly.
Check diaphragm and close
cover.
Check PEEP/P
MAX
line for
disconnect or leak.
Select Standby Mode and
switch back to the previous
ventilation mode.
!
(1)
VOLUME ALARMS OFF
Volume alarms disabled by
operator.
Priority Alarm Message
Probable Cause
Remedy
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...