12
Instructions for Use Fabius MRI SW 3.n
For Your Safety and that of Your Patients
WARNING
Always keep a breathing bag at hand. If venti-
lation of the patient is compromised, the
patient must be immediately ventilated with a
separate emergency ventilator.
WARNING
Remove any equipment mounted to the
machine before transport. The writing table
should also be free of all objects and pushed
back in the locked position.
If these precautions are not followed, the
device may tip over and pose a risk to safety.
WARNING
Do not place any object on this machine
unless it is specifically labeled to be used in
an MR scanning room and on a Fabius MRI
anesthesia system. Objects placed on this
machine that are not designed for use with
this anesthesia system may be strongly
attracted to the magnet and may cause seri-
ous injury or death when the machine is used
in an MR scanning room.
WARNING
All the procedures described in this manual
require that any use of ferromagnetic tools is
done only while the Fabius MRI is out of the
scanner room.
WARNING
Do not bring any ferromagnetic tools or equip-
ment into the scanning room.
Ferromagnetic objects (made of steel, iron, or
stainless steel) are strongly attracted to the
magnet and can become harmful projectiles.
WARNING
Be careful in handling the power cord and
main power plug. These parts still contain
minor magnetic components. The power cord
can be attracted to MRI system.
WARNING
Do not use any type of Desflurane vaporizer in
the MR environment. In an MR environment
functionality of the Desflurane vaporizer will
be compromised.
WARNING
A pre-use checkout procedure must be per-
formed immediately before each use of the
Fabius MRI. A recommended procedure is
provided in this Instruction for Use, see
“Appendix 1 - Daily and Pre-use Checkout
Form” on page 217.
WARNING
To avoid electrical shock hazard:
Due to the risk of electrical shock, do not
remove any component cover.
Refer any servicing to DrägerService.
Use only hospital-grade grounded electrical
outlets and power cord.
This device is to be used only in rooms with
line power installations complying with
national safety standards for hospital patient
rooms (e.g., IEC 60601-1 "Medical electrical
equipment").
Make sure the external equipment is hospital-
grade grounded (regarding national regula-
tions) before connecting the equipment.
Disconnect the power supply from the electri-
cal outlet before cleaning or servicing. Let it
dry completely before reconnecting it to the
electrical outlet.
Always ensure that the clamp for the power
cord at the power supply end is tight, thus pre-
venting an accidental disconnect from the
unit.
Do not connect additional external equipment
other than equipment specified by Dräger.
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...