Operation
116
Instructions for Use Fabius MRI SW 3.n
1
Remove the pipeline supply hoses from the
central supply.
z
Press the O
2
flush to depressurize the entire
system.
Moving the Machine
The Fabius MRI has a central caster brake bracket
on the front of the machine. To unlock the castors,
release the brake by lifting the bracket (e.g., with
the tip of your foot).
CAUTION
Do not attempt to move the anesthesia machine
while the casters are locked.
CAUTION
Watch hoses connected to the patient while mov-
ing the Fabius MRI.
CAUTION
Do not push or pull the Fabius MRI using the
absorber system, vaporizers, ventilator, or moni-
tor.
CAUTION
The Fabius MRI should be moved by people who
can handle its weight (approximately 425 lb
(192.8 kg)). Take extra care to avoid tipping when
moving on ramps, around corners, and over
raised thresholds such as door frames and eleva-
tors.
013
1
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...