Operation
106
Instructions for Use Fabius MRI SW 3.n
The parameters that can be set for SIMV/PS mode
are shown the adjacent table, along with their
adjustment ranges and factory default values.
z
To access the
Trigger
,
Insp Flow
,
T
INSP
, and
T
IP
:T
I
parameters, press the »
MORE
« key on
the SIMV/PS screen.
z
After the mode change is confirmed, the
»
SIMV/PS
« key LED switches from blinking to
constantly on, the ventilator switches to the
SIMV/PS mode, and the waveform is restored.
Ventilation Parameter
(SIMV/PS Mode)
Adjustment
Range
Factory
Default
Value
Pressure Limitation
P
MAX
[cmH
2
O] ([hPa])
15 to 70
min. PEEP+10
and
>
Δ
P
PS
+PEEP
40
Tidal Volume
V
T
[mL]
20 to 1100
600
Frequency
Freq [bpm] ([1/min])
4 to 60
12
Support pressure
Δ
P
PS
[cmH
2
O] ([hPa])
3 to 20, OFF
10
PEEP [cmH
2
O] ([hPa]) 0 to 20
0
Trigger Sensitivity
Trigger [L/min]
2 to 15
2
Inspiratory Flow
Insp Flow
[L/min]
10 to 85
30
SIMV insp. time
T
INSP
[seconds]
0.3 to 4.0
1.7
Insp pause time:
Insp time
T
IP
:T
I
[%]
0 to 50
10
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...