Instructions for Use Fabius MRI SW 3.n
73
Assembly
WARNING
For operation in magnetic fields, it is not per-
mitted to connect the vaporizer via hose con-
nectors or tapered connectors with the
anesthesia workstation.
WARNING
Risk of injury
Do not use ferromagnetic keyed filler or drain
adapters or tools when the filling or draining
procedure is carried out in magnetic fields.
Ferromagnetic adapters or tools can be
moved by magnetic attraction.
WARNING
Risk of injury
Vaporizers have to be mounted and removed
outside of the MR room.
Do not exchange any vaporizer in the MR
room.
Vaporizers can be moved by magnetic attrac-
tion.
WARNING
Risk of injury
Do not use filling adapters without a label
“MRI”. Adapters can be moved by magnetic
attraction.
CAUTION
Only Vapor 2000 vaporizers can by used on the
Fabius MRI in MRT scanner rooms.
CAUTION
Risk of injury
Dräger metal keyed filler adapters and Dräger fill
adapters are labeled with “MRI” are not ferromag-
netic.
Summary of Contents for Fabius
Page 36: ...36 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 84: ...84 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 130: ...130 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 192: ...192 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 215: ...Instructions for Use Fabius MRI SW 3 n 215 Technical Data 311 ...
Page 216: ...216 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 218: ...218 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 226: ...226 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 228: ...228 Instructions for Use Fabius MRI SW 3 n This page intentionally left blank ...
Page 231: ...Instructions for Use Fabius MRI SW 3 n 231 This page intentionally left blank ...